Home » KMUTT Library » PaperAdvance


Safety Distance for Medical Equipments Based on 2G and 3G Mobile Systems PDF Print E-mail
User Rating: / 0
PoorBest 
Written by zeezom   
Monday, 31 January 2011 11:27

Safety Distance for Medical Equipments Based on 2G and 3G Mobile Systems


Abstract_
In this study, two different mobile communication operators
provide services in Turkey have been choosen that each
operator has both 2G and 3G services. In this study,
electromagnetic interefernce distance to medical equipments
located in The Medical School Hospital of Akdeniz University
sourced from mobiled phones have been examined. Through
out different units in the hospital environment, 30 different
measurements carried out, and deterioration in audio and
visual signal reaction of devices was found to be associated
with the distance to mobile phones. Electromagnetic
interference, particularly of the ECG and ted EEG device
was observed, and exposure begins with range 1.25m
distance.

 

Add comment


Security code
Refresh